IRON LaboratoriesPJLA

ISO/IEC 17025:2005 Accredited

Marihuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 77802 *Test Certificate #: 116930-002
Client Name, Sample Details

Sugar Creek Bottling Co.
Indianapolis, IN 46220
Sample: 1200 mg #3
Type: Tincture
Method: FE04U

Test Conditions

Prepsheet ID#: MIP190514
Scale: XS205-MI2
Temp: 22.3 °C
Baro PE: 979.6 hPa
Analyst: KEB
Technician: KEB

Sample ID#: 116930

Harvest/Process Date: 05/14/2019
Serving Size (g): 1
Date Received: 05/14/2019
Test Date: 05/14/2019
Valid Through: 05/14/2020

Sample #116930
Sample #116930 Sample #116930
Test
Compounds
THC THCA CBD CBDA CBN CBG* CBC* THCV* CBDV* Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 0.0 N/D 4.1 N/D N/D 0.0 0.1 N/D 0.0 4.2 0.0 4.1 4.2
Amount (mg/g) 0.3 N/D 40.7 N/D N/D 0.3 0.5 N/D 0.2 42 0.3 40.7 42.0
Amount per Serving (mg) 0.3 N\D 40.7 N\D N\D 0.3 0.5 N\D 0.2 42.0 Serving Size~ (g): 1.0
LOQ (mg/g) 0.18 0.18 0.18 0.18 0.18 0.18 0.18 0.18 0.18   %Decarb. THC CBD
±%RPD 3.2 1.5 1.6 1.4 1.0 1.6 1.3 1.6 1.0 100% 100%

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45.

Total THC and CBD is the calculated sum of THC or CBD and the amount of THC or CBD derived from THCA or CBDA, respectively. These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value. Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, CBG, CBC, THCV, and CBDV.

%Decarb. THC and CBD refers to the percentage of THC or CBD relative to THCA or CBDA, respectively.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Test Certificate No. 116930-002 supersedes Test Certificate No. 116930-001

ACR - LM Iron laboratories ACR - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2005 Testing Laboratory laboratory, accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 1 of 2

IRON LaboratoriesPJLA

ISO/IEC 17025:2005 Accredited

QC Report

Testing Accreditation #: 77802 *Test Certificate #: 116930-002
Client Name, Sample Details

Sugar Creek Bottling Co.
Indianapolis, IN 46220
Sample: 1200 mg #3
Type: Tincture
Method: FE04U

Test Conditions

Prepsheet ID#: MIP190514
Scale: XS205-MI2
Temp: 22.3 °C
Baro PE: 979.6 hPa
Analyst: KEB
Technician: KEB

Sample ID#: 116930

Harvest/Process Date: 05/14/2019
Serving Size (g): 1
Date Received: 05/14/2019
Test Date: 05/14/2019
Valid Through: 05/14/2020

Sample #116930
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Cannabidivarin (CBDV) 0 5.053843493 5.072419004 100.37 5.122084774 101.35 0.97%
Cannabidiolic Acid (CBDA) 0 5.18208321 5.270890613 101.71 5.199107675 100.33 1.37%
Cannabigerol (CBG) 0 5.160904001 5.243947944 101.61 5.161483291 100.01 1.59%
Cannabidiol (CBD) 0 5.270103806 5.446064034 103.34 5.358711547 101.68 1.62%
Δ9-Tetrahydrocannabivarin (THCV) 0 5.102409165 5.158659134 101.10 5.240859198 102.71 1.58%
Cannabinol (CBN) 0 4.81760327 4.875050822 101.19 4.923708404 102.20 0.99%
Δ9-Tetrahydrocannabinol (THC) 0 5.254390697 5.247128322 99.86 5.416069637 103.08 3.17%
Cannabichromene (CBC) 0 4.706563352 4.815434414 102.31 4.877670472 103.64 1.28%
Tetrahydrocannabinolic acid (THCA) 0 4.75329088 4.837690044 101.78 4.910435869 103.31 1.49%

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Test Certificate No. 116930-002 supersedes Test Certificate No. 116930-001

ACR - LM Iron laboratories ACR - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2005 Testing Laboratory laboratory, accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 2